Anode and Cathode in Pacemakers
One concept that can cause confusion when discussing how pacemakers work is the difference between anode and cathode, especially regarding the flow of electric current and stimulation of the heart. To understand it properly, we need to review some basic principles of electricity and how they apply in the medical setting.

Direction of current: historical convention vs physical reality
Traditionally, electricity has been represented as a flow from the positive pole to the negative pole of a source (such as a battery). This representation is called conventional current and was adopted before the electron was discovered.
Physical reality, however, is different. Electrons, the particles that actually move and generate current, travel in the opposite direction: from the negative pole to the positive pole. This is the real electron current, which must be taken into account especially in practical applications such as cardiac pacing.
Application in pacemakers
In a pacemaker system, electric current travels from the generator (the source) to the heart (the load). This is where it is key to understand how electrons behave and how the lead poles are named.
- The pacemaker battery drives electrons from its negative pole.
- The electrons travel along the lead to the lead cathode, which is the negative pole of the lead implanted in the heart.
- From the cathode, the electrons leave the lead and enter the heart tissue, producing depolarisation and therefore myocardial contraction.
- Electrons that have passed through the heart tissue return to the pacemaker through the lead anode, which collects the current and directs it back to the positive pole of the battery, closing the circuit.

So, in a pacemaker, the lead cathode is the stimulating pole, the one that releases electrons into the tissue. The anode acts as the return point for the current.
Clinical importance
Correctly understanding the role of the anode and cathode in pacemaker leads is essential to interpret pacing configurations such as unipolar or bipolar, and to understand phenomena such as capture threshold or the direction in which the stimulus spreads through heart tissue.
This knowledge is also essential for correct device programming and for properly interpreting ECG recordings during follow-up of patients with pacemakers.