What Is the Pacemaker Threshold?
Pacemaker threshold is a fundamental concept in cardiac pacing. It refers to the minimum electrical stimulus needed to depolarise the myocardium and produce an effective contraction, that is, for the pacemaker to "capture" the heart (known as capture). Although people generally talk about the pacemaker threshold, we actually measure the lead threshold, and it has to be done for each lead if, for example, the pacemaker is dual-chamber.
How is the pacemaker threshold measured?
The threshold is assessed by observing the heart's response to different combinations of voltage and pulse duration. In clinical practice, the usual approach is to keep the pulse duration constant (for example 0.4 ms) and gradually reduce the voltage until the stimulus no longer produces a QRS.

The image above clearly shows this process:
- At 1.5 V, 1.25 V and 1.0 V with a duration of 0.4 ms, the pacemaker does capture the ventricle, as shown by a paced QRS after each spike.
- At 0.75 V with 0.4 ms there is still capture: this is the threshold, the minimum effective value.
- A 0.5 V, a QRS no longer follows the spike of the pacemaker. This indicates that the stimulus was not enough: there is no capture.
So in this example the threshold is: 0.75 V at 0.4 ms.
What is its clinical importance?
Knowing the pacing threshold is crucial to:
- Ensure effective pacing of the heart.
- Program an adequate safety margin (for example, output is usually programmed at twice the measured threshold).
- Optimise battery life of the pacemaker, avoiding using more energy than needed.
- Detect changes in the condition of the lead or heart tissue (for example oedema, fibrosis, lead displacement).
What if there is loss of capture?
If the stimulus does not produce a QRS (as seen in the last part of the image at 0.5 V), we speak of loss of capture. This can have many causes: a rise in threshold, poor lead position, lead fracture, interference, etc. The patient should therefore be reassessed and the device programming adjusted if necessary. However, modern pacemakers have algorithms that increase the output if the threshold rises.
Conclusion:
The pacemaker threshold defines the minimum energy needed to stimulate the heart effectively. Assessing and programming it correctly is key to safe, long-lasting device function. As the image shows, the moment capture disappears (no QRS after the spike) precisely marks this fundamental value.